New TopShim run:

version = topshim 2.1.0
date = 2025-01-08
time = 23:05:27.142
command = topshim plotall 1H o1p=2.014700 ss lockoff selwid=0.1

FINE SHIMMING

Parameters:

probe = Z154039_0001
solvent = None
shim nucleus = 1H
nucleus optimised for = 1H
o1p (set) = 2.01 ppm
selective excitation width = 0.1 ppm

Results:


1D Shimming:
Iteration 1: FWHM Before=16.8 Hz, Target= 0.2 Hz, After= 0.4 Hz
Iteration 2: FWHM Before= 0.4 Hz, Target= 0.2 Hz, After= 0.4 Hz
Iteration 3: FWHM Before= 0.4 Hz, Target= 0.2 Hz, After= 1.6 Hz
Iteration 4: FWHM Before= 1.6 Hz, Target= 0.2 Hz, After= 0.8 Hz
Iteration 5: FWHM Before= 0.8 Hz, Target= 0.3 Hz, After= 0.3 Hz
Iteration 6: FWHM Before= 0.3 Hz, Target= 0.4 Hz, After= 1.3 Hz
Iteration 7: FWHM Before= 1.3 Hz, Target= 0.2 Hz, After= 0.9 Hz
Iteration 8: FWHM Before= 0.9 Hz, Target= 0.2 Hz, After= 0.3 Hz
Iteration 9: FWHM Before= 0.3 Hz, Target= 0.3 Hz, After= 0.5 Hz
Iteration 10: FWHM Before= 0.5 Hz, Target= 0.2 Hz, After= 1.4 Hz
Slow convergence detected. Shimming was aborted.
Using the best result as the final shim state.

Lineshape quality factor (1D spectrum, non-spinning) = 0.44

shim changes: 
Z	-4
Z2	-1
Z3	-178
Z4	+331
Z5	-88
Z6	-895
Z7	-2315
Z8	+11883

duration = 2 min 6 sec
completed successfully