New TopShim run:

version = topshim 2.1.0
date = 2025-01-09
time = 10:04:03.979
command = topshim plotall 1H o1p=1.384200 ss lockoff selwid=0.1

FINE SHIMMING

Parameters:

probe = Z800701_0070
solvent = None
shim nucleus = 1H
nucleus optimised for = 1H
o1p (set) = 1.38 ppm
selective excitation width = 0.1 ppm

Results:


1D Shimming:
Iteration 1: FWHM Before= 0.8 Hz, Target= 0.2 Hz, After= 0.2 Hz
Iteration 2: FWHM Before= 0.2 Hz, Target= 0.2 Hz, After= 0.3 Hz
Iteration 3: FWHM Before= 0.3 Hz, Target= 0.2 Hz, After= 0.2 Hz
Iteration 4: FWHM Before= 0.2 Hz, Target= 0.2 Hz, After= 0.3 Hz
Iteration 5: FWHM Before= 0.3 Hz, Target= 0.2 Hz, After= 0.6 Hz
Iteration 6: FWHM Before= 0.6 Hz, Target= 0.1 Hz, After= 0.2 Hz
Slow convergence detected. Shimming was aborted.
Using the best result as the final shim state.

Lineshape quality factor (1D spectrum, non-spinning) = 0.259

shim changes: 
Z	-31
Z2	-5
Z3	-19
Z4	+350
Z5	-507
Z6	-1780

duration = 57 sec
completed successfully