New TopShim run:

version = topshim 2.0.0
date = 2024-12-13
time = 14:49:29.981
command = topshim 

FINE SHIMMING

Parameters:

probe = Z820201_0105
solvent = CDCl3
shim nucleus = 2H
nucleus optimised for = 1H
o1p (from lock) = 7.24 ppm

Results:


1D Shimming:
Iteration  1: FWHM Before= 3.0 Hz, Target= 0.4 Hz, After= 0.4 Hz
Iteration  2: FWHM Before= 0.4 Hz, Target= 0.2 Hz, After= 0.4 Hz
Iteration  3: FWHM Before= 0.4 Hz, Target= 0.3 Hz, After= 0.4 Hz
Iteration  4: FWHM Before= 0.4 Hz, Target= 0.3 Hz, After= 0.4 Hz
Iteration  5: FWHM Before= 0.4 Hz, Target= 0.3 Hz, After= 0.4 Hz
Iteration  6: FWHM Before= 0.4 Hz, Target= 0.3 Hz, After= 0.5 Hz
Using the best result as the final shim state.

Lineshape quality factor (1D spectrum, spinning) = 0.445

shim changes: 
Z	-116
Z2	+4
Z3	+22
Z4	-967
Z5	-189

duration = 1 min 23 sec
completed successfully
finished Fri Dec 13 14:50:51 2024


========================================================================