New TopShim run:

version = topshim 2.1.0
date = 2023-03-15
time = 18:13:58.302
command = topshim 

FINE SHIMMING

Parameters:

probe = Z154039_0001
solvent = CDCl3
shim nucleus = 2H
nucleus optimised for = 13C
o1p (from lock) = 7.24 ppm

Results:


1D Shimming:
echo time must be reduced
Iteration 1: FWHM Before= 1.1 Hz, Target= 0.3 Hz, After= 0.8 Hz
Iteration 2: FWHM Before= 0.8 Hz, Target= 0.2 Hz, After= 0.9 Hz
Iteration 3: FWHM Before= 0.9 Hz, Target= 0.3 Hz, After= 0.3 Hz
Iteration 4: FWHM Before= 0.3 Hz, Target= 0.3 Hz, After= 0.5 Hz
Iteration 5: FWHM Before= 0.5 Hz, Target= 0.2 Hz, After= 0.3 Hz
Iteration 6: FWHM Before= 0.3 Hz, Target= 0.2 Hz, After= 0.3 Hz
Iteration 7: FWHM Before= 0.3 Hz, Target= 0.2 Hz, After= 0.4 Hz
Slow convergence detected. Shimming was aborted.
Using the best result as the final shim state.

Lineshape quality factor (1D spectrum, non-spinning) = 0.344

shim changes: 
Z	-325
Z2	-83
Z3	-28
Z4	-191
Z5	-167
Z6	-724
Z7	+1615
Z8	+5737

duration = 2 min 17 sec
completed successfully
finished Wed Mar 15 18:16:13 2023


========================================================================